Emerson I-0021 Manuel D’Utilisation

Page de 238
5 - 20
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Calibration parameters
↓↓
Zero gases
↓↓
In the menu "Zero Gas Definition" you can set up the zero gas concentration for all ranges
of the current channel of an MLT analyzer or analyzer module. The concentration unit
("ppm", "ppb", "%" etc.) is determined by the setup of the current channel (see section
5.1.10 p. 5-54).
Zero gas concentration setup:
♦ 
Select the variable with the 
88 
 -key or the 
 -key.
♦ 
Select any digit with the 
 -key or the 
 -key and adjust a new value with
the 
↑↑
 -key or the 
↓↓
 -key
resp. select the whole parameter with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Confirm the new value with the 
88 
 -key or
cancel and go back to the last value with the F2 key.
Measure
Back...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- Zero Gas Definition --
Zero gas concentration (all ranges):
0.00  ppm
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters – Zero Gases