Emerson I-0021 Manuel D’Utilisation
5 - 22
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Parameters set up or changing to submenus:
♦
Select the line of variables, any menu line or the function line with
the
the
↓↓
-key or the
↑↑
-key.
♦
Select the variable, start the function or change to the submenu with
the
the
88
-key or the
→
→
-key.
♦
Select any digit with the
←
←
-key or the
→
→
-key and adjust a new value with
the
↑↑
-key or the
↓↓
-key
resp. select the whole parameter with the
↑↑
-key or the
↓↓
-key.
♦
Confirm the new value with the
88
-key or
cancel and go back to the last value with the F2 key.
Line of variables "Alarm delay":
The alarm message of any enabled parameter will start after the time in this line is over.
Options: 0, 0.1, ... , 30.0 s.
Options: 0, 0.1, ... , 30.0 s.
Function line "Acknowledge alarms !":
Starting this function will delete a running alarm message. You should do this after the
signal is back in the allowed range to reset the alarm message for new events.
signal is back in the allowed range to reset the alarm message for new events.
Note:
Via the F2 key you will reach the menu "Analyzer Channel Status". In that menu and its
corresponding submenus you will find the status report of the current channel for
corresponding submenus you will find the status report of the current channel for
♦
Failures
♦
Maintenance requests
♦
Function controls
♦
Measurements/Alarms
♦
Events
♦
Hours of operation
and the operational settings of the current channel like
♦
range settings and
♦
response time (t
90
-time).