Emerson I-0021 Manuel D’Utilisation

Page de 238
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 27
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Range parameters
↓↓
Begin and end of ranges
↓↓
In the menu "Begin and End of Ranges" you can set up the begin and end of range for
each range of the current channel. To change to the other channels of an MLT analyzer or
analyzer module, you have to press the F3 key.
Set up parameters:
♦ 
Select any line of variables with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Select the variable with the 
88 
 -key or the 
 -key.
♦ 
Select any digit with the 
 -key or the 
 -key and adjust a new value with
the 
↑↑
 -key or the 
↓↓
 -key
resp. select the whole value with the 
↑↑
 -key or the 
↓↓
 -key.
Allowed minimum/maximum range values:
Concentrations in the lines "Absolute, range lower/upper limit"
of the menu "Autoranging Control" (see p. 5-29).
♦ 
Confirm the new value with the 
88 
 -key or
cancel and go back to the last value with the F2 key.
Lines "Actual begin of range" / "Actual end of range":
In these lines you will find the begin and end of the current range.
Measure
Channel
Back...
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Begin and End of Ranges --
Begin of range -1:
0.00  ppm
Begin of range -3:
0.00  ppm
Begin of range -4:
0.00  ppm
End of range -1:
50.00  ppm
End of range -2:
250.0  ppm
End of range -4:
10000  ppm
Begin of range -2:
0.00  ppm
End of range -3:
2500  ppm
Actual begin of range:
0.00  ppm
Actual end of range:
250.00  ppm
5.1 Analyzer Module Setup