Emerson I-0021 Manuel D’Utilisation

Page de 238
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 29
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Range parameters
↓↓
Autoranging control
↓↓
In the menu "Autoranging Control" you can set up the parameters for the autoranging
control of the current channel. If you want to set up the autoranging control of the other
channels of an MLT analyzer or analyzer module, you have to press the F3 key.
If the autoranging is enabled, the best range for the current concentration will be selected,
automatically.
Conditions:
♦ 
♦ 
The begin of range has to be zero for all four ranges of the current channel!
To setup: See menu "Begin and End of Ranges", p. 5-27.
♦ 
All four response times of the current channel must have the same value!
To setup: See menu "Response Times", previous page.
♦ 
The span gas concentration has to be in the correct range!
To setup: See menu "Span Gas Definition", section 5.1.1 p. 5-6.
The last three lines of the menu:
The parameters in these lines are factory settings and cannot be changed in this menu.
♦ 
Absolute, range lower limit: Normally is this value zero. It will only differ from zero for
differential measurements (see 5.1.12 p.5-57).
♦ 
Absolute, range upper limit: It is 120% of the largest end of range from putting into
operation. Higher values will not be accepted as end of range values!
Measure
Channel
Back...
More...
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Autoranging Control --
Switch level hysteresis:
10  %
Usage of range -2:
Enabled
Usage of range -3:
Enabled
Usage of range -4:
Enabled
Absolute, range lower limit:
0.00  ppm
Calculate switch levels automatically:
Enabled
Usage of range -1:
Enabled
Absolute, range upper limit:
1000000  ppm
5.1 Analyzer Module Setup
5.1.3 Range parameters – Autoranging control