Emerson I-0021 Manuel D’Utilisation

Page de 238
5 - 72
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Serial interface setup:
Press the 
88 
 -key or the 
 -key in the line "Serial interface setup..." of the menu "Local
SIO Configuration Parameters" to change to the corresponding submenu:
In the submenu "Serial Interface" you can determine the parameters of the data transfer
between the analyzer module and external implements. The setups in this menu depend
on the configuration of the analyzer module and the corresponding implement. The
specification of the serial interface is described in another manual.
Set up parameters:
♦ 
Select any line of variables or the menu line with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Select the variable or change to the submenu with the 
88 
 -key or the 
 -key.
♦ 
Select any digit with the 
 -key or the 
 -key and adjust a new value with
the 
↑↑
 -key or the 
↓↓
 -key
resp. select the whole parameter with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Confirm the new value with the 
88 
 -key or
cancel and go back to the last value with the F2 key.
Options:
Baudrate:
300
1200
2400
4800 9600 19200
Data bits:
7
8
Stop bits:
1
2
Parity:
None
Even
Odd
Echo mode:
Enabled
Disabled
Handshake:
None
Xon/Xoff
Transmission delay:
0 ... 100
Type of installed
serial interface:
RS 232 RS 485/2w RS 485/4w RS 485/4w-Bus None
Communication protocol:
AK
MODBUS RTU
None
Note: Via the line "Special protocol definitions..." you can change to a submenu, where
you can set up the parameters of the "AK" or "MODBUS RTU" communication protocol.
Measure
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Serial Interface ---
Baudrate:
19200
Stop bits:
1
Parity:
None
Echo mode:
Disabled
Handshake:
Xon/Xoff
Data bits:
8
Type of installed serial interface:
RS232
Back...
Transmission delay:
0
Communication protocol:
AK
Special protocol definitions...