Emerson I-0021 Manuel D’Utilisation
5 - 74
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Inputs and outputs
↓↓
DIO-module(s)
↓↓
In the menu "Local DIO Boards Setup" you can set up the configuration of the local DIO
modules of the analyzer module. Each DIO module is composed of 8 digital inputs and 24
digital outputs. You can attach a function to each input (e.g. opening of a valve) and a
signal to each output (e.g zero valve) by a corresponding code.
modules of the analyzer module. Each DIO module is composed of 8 digital inputs and 24
digital outputs. You can attach a function to each input (e.g. opening of a valve) and a
signal to each output (e.g zero valve) by a corresponding code.
See the tables on the following pages:
+
+
Set up parameters:
♦
Select the line of variables or the function line with the
↓↓
-key or the
↑↑
-key.
♦
Select the variable or start the function with the
88
-key or the
→
→
-key.
♦
If asked after function start: Confirm with the F2 key (Yes) or
cancel and go back to the menu page with the F4 key (Back...) or the
cancel and go back to the menu page with the F4 key (Back...) or the
←
←
-key.
♦
Variable changing:
Select any digit with the
Select any digit with the
←
←
-key or the
→
→
-key and adjust a new value with
the
↑↑
-key or the
↓↓
-key.
♦
Confirm the new value with the
88
-key or
cancel and go back to the last value with the F2 key.
Measure
Back...
NEXT_DIO
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Local DIO Boards Setup --
Number of selected input:
1
Number of selected output:
1
- Signal code:
111
Reset output fail !
Board status:
Normal
Current status of selected input:
Low
Current status of selected output:
High
- Function code:
13
Slot number of selected board:
6
5.1
Analyzer Module Setup
5.1.18 Inputs and Outputs – Local DIO
[2nd menu page via F5 (More...)]