Tektronix TVS600 & TVS600A Series Manuel D’Utilisation

Page de 410
ROSCillator Subsystem
2–150
TVS600 & TVS600A Command Reference
ROSCillator:SOURce
ROSCillator:SOURce?
Sets or queries the source of the 10 MHz clock reference for the acquisition
system. The choices are CLK10 which specifies the VXIbus 10 MHz reference,
and  INTernal which specifies the internal waveform analyzer oscillator.
[SENSe:]ROSCillator:SOURce <source>
[SENSe:]ROSCillator:SOURce?
<source>
Query response
CLK10
INTernal
CLK10
INT
INT
Execution Error –141, “Invalid character data”
Attempted to set source to an illegal value.
None
Command:
ROSC:SOUR CLK10
Query:
ROSC:SOUR?
Response:
CLK10
SWEep:TINTerval
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands