Tektronix TVS600 & TVS600A Series Manuel D’Utilisation

Page de 410
TRIGger:B Subsystem
TVS600 & TVS600A Command Reference
2–251
TRIGger:B:FILTer[:LPASs]
TRIGger:B:FILTer[:LPASs]?
Sets or queries the state of the 50 kHz, low-pass filter for the B trigger circuit.
Components of the trigger signal above 50 kHz are attenuated when the LPASs
filter is on. It may only be used when DC coupled and with TRIGger:SOURce
set to INTernal. Only one trigger filter (i.e., LPAS, HPAS, or NREJ) may be
enabled at a time.
The INPut:FILTer is separate from the trigger filter and is located before the
trigger pickoff in the signal path.
TRIGger:B:FILTer[:LPASs][:STATe] <boolean>
TRIGger:B:FILTer[:LPASs][:STATe]?
<boolean>
Query response
<NRf >
1 or ON
0 or OFF
<NR1>
1
0
0
None
Enabling the low pass filter will set TRIGger:B:COUPling to DC and set
TRIGger:B:FILTer:NREJect to OFF.
When TRIGger:A:SOURce and TRIGger:B:SOURce are set to the same signal,
changing LPAS for one will change LPAS for the other.
Command:
TRIG:B:FILT ON
Query:
TRIG:B:FILT?
Response:
1
TRIGger:B:COUPling
TRIGger:B:FILTer:HPASs
INPut:FILTer
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands