Tektronix TVS600 & TVS600A Series Manuel D’Utilisation
TRIGger:LOGic Subsystem
TVS600 & TVS600A Command Reference
2–263
TRIGger:LOGic:FUNCtion
TRIGger:LOGic:FUNCtion?
TRIGger:LOGic:FUNCtion?
TVS600A Models Only
Sets or queries the logical combination of the input channels for the main logic
trigger. When TRIGger:LOGic:CLASs is set to PATTern, this command applies
to channel 1, 2, 3 and 4. When TRIGger:LOGic:CLASs is set to STATe, only
channels 1, 2 and 3 are logically combined.
trigger. When TRIGger:LOGic:CLASs is set to PATTern, this command applies
to channel 1, 2, 3 and 4. When TRIGger:LOGic:CLASs is set to STATe, only
channels 1, 2 and 3 are logically combined.
:A is a pre–defined alias for :SEQuence[1].
This command only has effect when TRIGger:TYPE is set to LOGic.
TRIGger[:A]:LOGic:FUNCtion <logic_func>
TRIGger[:A]:LOGic:FUNCtion?
<logic_fun>
Query response
AND
NAND
NOR
OR
NAND
NOR
OR
AND
NAND
NOR
OR
NAND
NOR
OR
None
AND
Command Error –141, “Invalid character data”
Attempted to set FUNCtion to an illegal value.
Attempted to set FUNCtion to an illegal value.
None
Command:
TRIG:LOG:FUNC NOR
Query:
TRIG:LOG:FUNC?
Response:
NOR
TRIGger:TYPE
TRIGger:LOGic:CLASs
TRIGger:LOGic:CLASs
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands