TRIGger:LOGic Subsystem
TVS600 & TVS600A Command Reference
2–261
TRIGger:TYPE
TRIGger:LOGic:CONDition
TRIGger:LOGic:CONDition?
TRIGger:LOGic:CONDition?
TVS600A Models Only
Sets or queries the trigger condition for each channel when logic triggering is
selected. The pattern argument refers to the two or four channels of the instru-
ment in a left to right syntax for channel 4(2) through channel 1. A “1” indicates
a High, a “0” indicates a Low, and an “X” indicates a Don’t Care is selected.
selected. The pattern argument refers to the two or four channels of the instru-
ment in a left to right syntax for channel 4(2) through channel 1. A “1” indicates
a High, a “0” indicates a Low, and an “X” indicates a Don’t Care is selected.
:A is a predefined alias for :SEQuence[1].
This command only has effect when TRIGger:TYPE is set to LOGic.
The condition of channel 4 is ignored when TRIGger:LOGic:CLASs is set to
STATe.
STATe.
The condition of Bits 3 and 4 are ignored when the instrument is configured with
two channels.
two channels.
TRIGger[:A]:LOGic:CONDition <pattern>
TRIGger[:A]:LOGic:CONDition?
<paggern>
Query response
LCXXX0
LCXXX1
....
LCM1111
LCXXX1
....
LCM1111
LCXXX0
LCXXX1
LCXXX1
LCM1111
None
LC0000
Command Error –141, “Invalid character data”
Attempted to set either condition parameter to an illegal value.
Attempted to set either condition parameter to an illegal value.
None
Command:
TRIG:LOG:COND LC1100
Related Commands
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples