Tektronix TVS600 & TVS600A Series Manuel D’Utilisation
TRIGger:TRANsition Subsystem
TVS600 & TVS600A Command Reference
2–299
TRIGger:TRANsition:SOURce
TRIGger:TRANsition:SOURce?
TRIGger:TRANsition:SOURce?
TVS600A Models Only
Sets or queries the trigger source for all transition triggering classes.
:A is a predefined alias for :SEQuence[1].
This command only has effect when TRIGger:TYPE is TRANsition.
Setting source to INTernal1 causes the signal from INPut1 and [SENSe:]VOLT-
age1 to be the trigger source.
age1 to be the trigger source.
TRIGger[:A]:TRANsition:SOURce <source>
TRIGger[:A]:TRANsition:SOURce?
<source>
Query response
INTernal
INTernal1
INTernal2
INTernal3
INTernal4
INTernal1
INTernal2
INTernal3
INTernal4
INT
INT1
INT2
INT3
INT4
INT1
INT2
INT3
INT4
None
INT
Command Error –141, “Invalid character data”
Attempted to set source to an illegal value.
Attempted to set source to an illegal value.
None
Command:
TRIG:TRAN:SOUR INT3
Query:
TRIG:TRAN:SOUR?
Response:
INT3
TRIGger:TYPE
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands