*TST? (Query Only)
Initiates an internal self-test and returns a failure code. If more than one failure
occurred, only the test number of the first failure is reported. A value of zero
indicates there were no failures. This query is only valid when looping is
disabled. Detailed test results are available with the TEST:RESult:VERBose?
command.
*TST? and TEST? perform the same function.
<failure_code>
Query response
Not applicable
<NR1>
0 (no 
failure)
1000 – 1999
(Self-test failure)
2000 – 2999
(Calibration failure)
Not applicable
Execution Error –221, “Settings conflict”
Executed *TST? when TEST:CONTrol:LOOP is ON.
None
Query:
Response:
TEST
TEST:RESults:VERBose?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples