Tektronix TVS600 & TVS600A Series Manuel D’Utilisation
CALCulate Subsystem
2–36
TVS600 & TVS600A Command Reference
CALCulate:FILTer:FREQuency:LPASs
CALCulate:FILTer:FREQuency:LPASs?
CALCulate:FILTer:FREQuency:LPASs?
Sets or queries the limit frequency (<cutoff >) above which the filter attenuates
all frequency components. The :LPASs frequency is ignored when filter type is
set to bandpass, notch or high pass. Use the command CALCulate:FILTer:FRE-
Quency LPASs to enable the low-pass filter.
all frequency components. The :LPASs frequency is ignored when filter type is
set to bandpass, notch or high pass. Use the command CALCulate:FILTer:FRE-
Quency LPASs to enable the low-pass filter.
CALCulate<n>:FILTer[:GATE]:FREQuency:LPASs <cutoffā>
CALCulate<n>:FILTer[:GATE]:FREQuency:LPASs?
<n>
1
Query response
1
2
3
4
2
3
4
Not applicable
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
the default is calculate block 1.
<cutoff> (Limit frequency)
1
Query response
<NRf >
<NR3>
1
The default multiplier for the <cutoff > parameter is HZ for hertz. You can also use the
multipliers KHZ for kilohertz, MHZ for megahertz, and GHZ for gigahertz
multipliers KHZ for kilohertz, MHZ for megahertz, and GHZ for gigahertz
250.0E+6
Execution Error –222, “Data out of range”
Attempted to set the low-pass cutoff frequency to an illegal value.
Attempted to set the low-pass cutoff frequency to an illegal value.
None
Command:
CALC1:FILT:FREQ:LPAS 10E6
Query:
CALC1:FILT:FREQ:LPAS?
Response:
10.0E+6
CALCulate:FILTer:FREQuency:HPASs
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands