As can be seen from the above reactions, the calculation of the ammonia 
concentration and overall calibration of the instrument must be done 
carefully, if accurate ammonia concentrations are to be measured. 
1.2  S
PECIAL 
C
ONSIDERATIONS 
F
OR 
A
MMONIA 
M
EASUREMENT
 
Ammonia is a difficult gas to measure due to its chemical characteristics. 
The gas tends to adsorb onto surfaces and diffuse into many materials. The 
following precautions should be observed when designing ammonia 
sampling systems and connecting them to the T201 analyzer: 
 
Do NOT use copper tubing or fittings designed for household plumbing. 
 
Use ONLY Chromatography grade (cleaned, passivated) stainless steel 
tubing.  
 
Use ONLY glass for sample inlet manifold. 
 
ALWAYS keep the tubing from the M702SS calibrator to the analyzer 
as short as possible. USE stainless steel tubing throughout, especially 
from the ammonia calibration bottle to the calibrator. 
 
If possible HEAT the sample line and DRY the sample gas. 
 
Sample filters will cause response delays on both the rise and fall of the 
sample signal. 
 
These rules apply also to your calibrator’s internal plumbing.  It is highly 
recommended that you use a M702SS calibrator in conjunction with the 
T201. The calibrator has stainless steel plumbing throughout. The T700 
calibrator contains Teflon tubing and internal MFC’s that can be affected by 
the ammonia gas, therefore, the T700 calibrator is not recommend for use 
with the T201 analyzer. It is important to keep the sampling system well 
maintained. 
1.3 
 
S
AMPLE 
F
ILTRATION
 
The instrument can be provided with an optional stainless steel sample 
filter P/N 05571. For minimum response time operation, the instrument 
should be supplied with clean sample gas at ambient pressure. 
An additional filter resides in the T201 pneumatic pathway. The filter is a 
¼” diameter stainless steel sintered filter used to protect the reaction cell 
flow control orifice from plugging. This filter will plug rapidly if 
contaminated sample gas is not filtered before entering the analyzer. 
Note 
In situations where the sample gas is known to be contaminated with 
particulate material, it is advisably to add extra filtration. However, it must be 
noted that the response time is directly extended through filter additions.