Calibration Procedure 
Teledyne API Model T201 NH
Analyzer Operator Manual 
16 
07271B DCN6646 
  
2.1  Z
EROING THE 
A
NALYZER
 
The analyzer can be zeroed by either applying zero air straight to the 
sample port or through the optional Zero/Span valves, if installed. It is 
important to remember that if the analyzer was previously sampling 
ammonia gas, prior to conducting a Zero calibration, it will take some 
period of time before the zero baseline is reached. The process of zeroing 
the analyzer consists of delivering dry zero air to the analyzer. The 
operator then manually zeros the TN
x
 and NO
x
 channels individually 
through the user interface. The following Tables outline the steps 
necessary to zero the analyzer.  
Table 2-1. Zero Calibration Procedure – Zero Gas through the SAMPLE Port 
Important: Each channel (TN
x
 and NO
x
) must be zeroed individually 
Step No. 
Action 
Comment 
1. Press 
CAL 
The T201 enters the calibrate mode from sample 
mode. Confirm zero gas is flowing past the sample 
port. 
2. Channel 
Selection 
Press 
TN
X
 or NO
x
 then Press ENTR
3. Range 
Selection 
Press 
LOW or HIGH range then Press ENTR.  
Always calibrate the LOW range, section 3.0 below. 
4. NH
3
 STB Reading  Wait for the displayed stability reading to fall below
1.0 PPB. The lower the stability reading the better the
zero value.  
5. Press 
ZERO 
If you change your mind after pressing ZERO, you 
can still press EXIT without zeroing the instrument.
You don’t have to enter concentration values. The
analyzer knows to apply 0 PPB. 
6. Press 
ENTR Pressing 
ENTR actually adjusts the Offset value for 
the selected channel. 
7. Press 
EXIT 
The T201 returns to sampling mode. 
8. Check 
Concentrations 
/Offsets 
All displayed concentrations should read zero. The 
offset values should be close to zero (-20 to 150 mv)