Analyzer Operator Manual
 
Calibration Procedure
 
07271B DCN6646 
17 
Table 2-2. Zero Calibration Procedure - Zero Gas through ZERO Port 
Step No. 
Action 
Comment 
1. Press 
CALZ 
The T201 enters the calibrate mode from sample 
mode. The zero gas is supplied through the ZERO
inlet port on the rear panel of the M501 NH
3
Table 2.1 
 
Notes 
  Since the zero gas concentration is defined as 0 ppb, it is not necessary 
to enter the expected concentration values.  
  Both the TN
x
 and NO
x
 channels must be zeroed individually. When you 
zero the NO
x
 channel both the offsets for NO
x
 and NO are configured. 
All readings should display zero following this action.  
  It is recommended that the calibration of both the TN
and NO
x
 channels 
be done at one time.  
  Always confirm that excess zero air is flowing past the sample or zero 
port, if the Zero/Span option is present, on the rear of the M501 NH
3
 
converter. There should always be some excess flow. If insufficient flow 
is supplied to the analyzer ambient gas will be entrained and the zero 
offset values will be incorrect. However, it is important not to over 
pressurize the analyzer.  
  It is a good idea to allow the analyzer to sample zero gas for extended 
periods following zero calibration. Large drifts can signify leaks or 
indicate an issue with the zero air source. 
 
2.2  S
PANNING THE 
A
NALYZER WITH 
N
ITRIC 
O
XIDE 
G
AS
 
The analyzer can be spanned by either applying nitric oxide gas straight to 
the Sample port or through the Zero/Span valve option, if installed. It is 
important to remember that if the analyzer was previously sampling
 
ammonia gas, prior to conducting a nitric oxide span calibration, it will 
take some period of time before the TN
x
 channel stabilizes. The process of 
spanning the analyzer consists of diluting bottled gas, using a calibrator 
and delivering to the analyzer. The operator then manually spans the TN
x
 
and NO
x
 channels individually through the user interface. The following 
Tables outline the manual steps necessary to span the analyzer.