4.0 MAINTENANCE 
SCHEDULE 
The maintenance requirements of the T201 are the same as a standard 
T200 NO
x
 analyzer but with additional maintenance items (Table 4-1). 
Please refer to Section 13 in the T200 Operator Manual for the T200 
maintenance schedule and for maintenance and repair procedures. 
The T201 requires the following additional maintenance items. 
Table 4-1. Preventative Maintenance Schedule 
 
Item 
Maintenance Interval 
Reference Section 
NO
2  
molybdenum converter 
Test every three months 
Section 8.2.3 T200 Manual 
M501 NH
3
 Converter 
Test every three months 
Replace annually or as necessary 
Reaction Cell 
Clean annually or as necessary 
Section 9.3.9 T200 Manual 
AutoZero flow check 
Quarterly as needed 
 
 
4.1  M501
 
NH
3
 
M
AINTENANCE 
 
The M501 NH
3
 external converter is operated at 825 ºCelsius. Because of 
this the stainless steel ¼” tube internal to the converter slowly oxidizes 
and may require replacement. Over time the converter’s efficiency will 
degrade, due to aging of the internal catalyst. This is characterized by a 
CONV value on the TNx channel of < 0.8 or > 1.2, despite the converter 
being at temperature. The efficiency of the converter should be checked 
quarterly. The converter should be cleaned out on a yearly basis by 
removing the outer tube and shaking the fine dust particles from both the 
outer and inner tubes. This is also a good time to replace the catalyst 
screen, if necessary. 
The following procedure describes how to disassemble and replace 
converter parts.  The various parts in the converter become delicate and 
brittle after prolonged exposure to high temperatures.  It is therefore a 
good idea to have a complete set of replacement parts on hand before 
starting, as listed in Table 4-2.