Teledyne T201 Manuel D’Utilisation

Page de 44
Teledyne API Model T201 NH
Analyzer Operator Manual
 
Maintenance Schedule
 
07271B DCN6646 
29 
4.3  R
EPLACING THE 
T
HERMOCOUPLE
 
The following instructions provide the necessary information to replace 
the existing thermocouple. Figure 4-2 is provided for reference. 
You must obtain the following replacement parts kit from TAPI: 
 
010820000 , Type K thermocouple 
You will need the following tools: 
 
Nutdriver, 
5
/
16
 
 
Nutdriver, 
11
/
32
 
 
Diagonal Cutter 
 
Philips head Screwdriver #2 
 
 
CAUTION 
Avoid damage to the unit: use only Type K thermocouple in the 
M501 NH
3
 converter. It can be distinguished by its wire colors: red 
and yellow. If you have any other thermocouple, do NOT install it; 
call TAPI to obtain the correct part. 
 
 
Figure 4-2. Thermocouple Location