SECTION 5—OPERATING INSTRUCTIONS
23
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm-up 
time (approximately 30 min.) while waiting for the O
2
 purity to 
reach maximum.
When the unit is turned on, the GREEN light will come on 
(SYSTEM OK/O
2
 greater than 85%). After five minutes, the 
oxygen sensor will be operating normally and will control the 
indicator lights depending on oxygen concentration values. 
Explanation of Oxygen Purity Indicator 
Lights - Model IRC5PO2
NOTE: For this procedure, refer to FIGURE 5.6.
GREEN light (O
2
) - Normal Operation.
YELLOW light (
) - Call supplier IMMEDIATELY. You may 
continue to use the concentrator unless instructed otherwise 
by your supplier. Be certain that backup oxygen is nearby.
RED light ( ) - Total unit shutdown. Switch IMMEDIATELY 
to backup oxygen supply. Call supplier IMMEDIATELY.
GREEN light - with YELLOW light flashing - Call supplier 
IMMEDIATELY. Oxygen sensor malfunctioning; you may 
continue to use the concentrator. 
O
2
Auto Shut Down 
Use Backup
Below 
Normal
Normal
RED
Call Supplier
YELLOW
GREEN
OXYGEN PURITY