Omega OMB-DAQBOARD-3000 Manuel D’Utilisation

Page de 108
5-10   Counter Input Modes  
918494 
DaqBoard/3000 Series User’s Manual  
 
 
 
 
Upper 16-bits of the 32-bit counter 
 
Lower 16-bits of the 32-bit counter 
Range (Hz) 
Ticksize (nS) 
Averaging 
Option 
 
Range (Hz) 
Ticksize (nS) 
Averaging 
Option 
15u – 1500u 
20833.333 
 
1 – 100 
20833.333 
150u – 15m 
2083.333 
 
10 – 1k 
2083.333 
1500u – 150m 
208.333 
 
100 – 10k 
208.333 
15m – 1500m 
20.833 
 
1k – 100k 
20.833 
150m – 15 
20.833 
10 
 
10k – 1M 
20.833 
10 
1500m – 150 
20.833 
100 
 
100k – 5M 
20.833 
100 
15 – 1500 
20.833 
1000 
 
1M – 5M 
20.833 
1000 
Frequency Ranges for a 16-bit value, sampling error is less than 0.21% 
Each frequency range given in the previous table-set can be exceded.  If the input waveform goes under-
range by too much, the counter value will top out at 65535 indicating you have reached the lowest possible 
frequency that can be measured on that range.  If the input waveform goes over range by too much, the 
counter will return values that are very course and have a lot of sampling error.  The values returned will 
have a small number of counts for the period duration.  If an input waveform cannot fit within one of the 
16-bit ranges shown above or requires much higher accuracy, then a 32-bit range should be considered.  
 
Full 32-bit Counter  
Range (Hz) 
Ticksize (nS) 
Averaging Option 
15u – 100 
20833.333 
150u – 1k 
2083.333 
1.5m – 10k 
208.333 
15m – 100k 
20.833 
150m – 1M 
20.833 
10 
1.5 – 5M 
20.833 
100 
15 – 5M 
20.833 
1000 
Frequency Ranges for a 32-bit Value, 
Sampling Error is Less than 0.21% 
 
The 32-bit ranges shown above are much wider than the 16-bit ranges, but also require the full 32-bit value 
to be returned.  Since digital or counter channels do not take up any time in the scan period there is no 
disadvantage in reading a 32-bit counter versus a 16-bit counter.   The 32-bit frequency ranges can also be 
exceeded with a loss of accuracy or topping out at 4,294,967,295 counts. 
Some measurements will require the accuracy of an input waveform to be free of sampling error, having only 
the absolute accuracy of the internal timebase as the source of error.  Sampling error can be averaged out to 
give the required result.  In most cases, the 3000 Series board can perform the required averaging on the 
values before they are returned to the PC.  The frequency ranges shown below will give a sampling error 
that is less than 10ppm or 1ppm. 
 
Full 32-bit Counter  
<10 ppm 
Full 32-bit Counter  
<1 ppm 
Range (Hz) 
Ticksize (nS) 
Averaging 
Option 
Range (Hz) 
Ticksize (nS) 
Averaging 
Option 
15u – 500m 
20833.333 
15u – 50m 
20833.333 
150u – 5 
2083.333 
150u – 500m 
2083.333 
1.5m – 50 
208.333 
1.5m – 5 
208.333 
15m – 500 
20.833 
15m – 50 
20.833 
150m – 5k 
20.833 
10 
150m – 500 
20.833 
10 
1.5 – 50k 
20.833 
100 
1.5 – 5k 
20.833 
100 
15 – 500k 
20.833 
1000 
 
15 – 50k 
20.833 
1000 
High Accuracy Frequency Ranges for a 32-bit Value 
that has little sampling error (<10ppm, <1ppm) 
 
If the input frequency is required to have less than 1 ppm sampling error and is greater than 50kHz, use the 
15– 50kHz, 1ppm range.  The values returned will have sampling error that is greater than 1ppm but they 
can be averaged by the PC software to further reduce the sampling error.