5-12   Counter Input Modes  
918494 
DaqBoard/3000 Series User’s Manual  
 
 
PULSEWIDTH:  OPT2:  Determines whether the pulsewidth is to be measured with a 16-bit (Counter 
Low), or 32-bit (counter High) counter.  Since pulsewidth measurements always have the “stop at the top” 
option enabled, this option dictates whether the measurement has a range of 0 to 65535 ticks, or 0 to 
4,294,967,295 ticks.  
PULSEWIDTH:  OPT4:  Allows the mapped channel to gate the counter.  When the mapped channel 
is high, the counter is enabled to count.  When the mapped channel is low, the counter is disabled, but 
holds the count value.   The mapped channel can be any other input channel. 
PULSEWIDTH:  OPT6:  This allows the mapped channel’s pulsewidth to be measured instead of the 
input channel.  The mapped channel can be any other input channel (post debounce). This option allows 
the counter to be used with any other input channel (post-debounce).  If the channel’s input is used 
elsewhere, for example, gating another counter, the counter for this channel does not need to go unused.  
Pulsewidth and Timing mode Accuracy  
The 3000 Series board has the ability to measure the pulsewidth of an input and the time between any two 
edges on any two inputs.  The time ranges are similar to those shown for period mode except that averaging 
is not available.  The ranges given below reflect this.   
 
Upper 16-bits of the 32-bit Counter 
 
Lower 16-bits of the 32-bit Counter 
Range (S) 
Ticksize (nS) 
Averaging 
Option 
 
Range (S) 
Ticksize (nS) 
Averaging 
Option 
800 – 80000 
20833.333 
 
10m – 1 
20833.333 
80 – 8000 
2083.333 
 
1m – 100m 
2083.333 
8 – 800 
208.333 
 
100u – 10m 
208.333 
800m – 80 
20.833 
 
10u – 1m 
20.833 
Pulsewidth and Time Ranges for a 16-bit Value 
Sampling error is less than 0.21% 
 
 
Full 32-bit Counter  
Range (S) 
Ticksize (nS) 
Averaging 
Option 
10m – 80000 
20833.333 
1m – 8000 
2083.333 
100u – 800 
208.333 
10u - 80 
20.833 
Pulsewidth and Time Ranges for a 32-bit Value 
Sampling error is less than 0.21% 
 
 
 
Full 32-bit Counter     <10 ppm 
Full 32-bit Counter     <1 ppm 
Range (S) 
Ticksize (nS) 
Averaging 
Option 
Range (S) 
Ticksize (nS) 
Averaging 
Option 
2 – 80000 
20833.333 
20 – 80000 
20833.333 
200m – 8000 
2083.333 
2 – 8000 
2083.333 
20m – 800 
208.333 
200m – 800 
208.333 
2m – 80 
20.833 
 
20m – 80 
20.833 
High Accuracy Pulsewidth and Time Ranges for a 32-bit Value 
 that has little sampling error (<10ppm, <1ppm)