MANGANESE HIGH RANGE
MANGANESE HIGH RANGE
MANGANESE HIGH RANGE
MANGANESE HIGH RANGE
MANGANESE HIGH RANGE
10 mL
SPECIFICATIONS
Range
0.0 to 20.0 mg/L
Resolution
0.1 mg/L
Accuracy
±0.2 mg/L ±3% of reading
Typical EMC
±0.1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 525 nm
Method
Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
th
edition,
 Periodate method. The reaction between
manganese and reagents causes a pink tint in
the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93709A-0
Citrate
1 packet
HI 93709B-0
Sodium periodate
1 packet
REAGENT SETS
HI 93709-01 Reagents for 100 tests
HI 93709-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Manganese HR on
the secondary LCD by pressing PROGRAM W and V.
• Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
Manganese HR