Hanna Instruments c 99 Manuel D’Utilisation

Page de 148
91
MANGANESE LOW RANGE
MANGANESE LOW RANGE
MANGANESE LOW RANGE
MANGANESE LOW RANGE
MANGANESE LOW RANGE
SPECIFICATIONS
Range
0 to 300 µg/L
Resolution
1 µg/L
Accuracy
±2 µg/L ±3% of reading
Typical EMC
±1 µg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 575 nm
Method
Adaptation of the PAN Method. The reaction
between manganese and the reagents causes an
orange tint in the sample.
REQUIRED REAGENT
Code
Description
Quantity
HI 93748A-0
Ascorbic acid
2 packets
HI 93748B-0
Alkaline-cyanide sol.
0.40 mL
HI 93748C-0
0.1% PAN indicator
2 mL
HI 93703-51
Dispersing Agent
4-6 drops (only when
necessary, see note)
REAGENT SETS
HI 93748-01 Reagents for 50 tests
HI 93748-03 Reagents for 150 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Manganese LR on
the secondary LCD by pressing PROGRAM W and V.
• Fill one cuvet with 10 mL of deionized water
(up to the mark).
• Fill a second cuvet with 10 mL of sample (up to
the mark).
• Add the content of one packet
of HI 93748A Ascorbic acid to
each cuvet, replace the caps
and shake gently until dissolution
is complete.
Manganese LR
10 mL
10 mL