8.2.3  P1.3: Selection of Automatic or Manual Temperature Compensation 
This feature lets you select between Automatic Temperature Compensation (ATC) 
and Manual Temperature Compensation (MTC). Meter default is ATC.  
From measurement mode 
Press SETUP key to enter Set Up menu.  
2  Press the HOLD/ENTER keys until you view 
parameter P1.3.  
Press the HOLD/ENTER key again. The upper display 
shows “ATC” and the lower display shows “YES” or 
“NO”.  
4 Press 
MI/▲ or MR/▼ keys to select between 
Automatic and Manual Temperature Compensation.  
•  “YES” activates the Automatic Temperature 
Compensation function. 
•  “NO” activates the Manual Temperature 
Compensation function. 
Press the HOLD/ENTER key to confirm selection and 
to return to the subgroup menu.  
6  Press the CAL/MEAS key to return to measurement 
mode.  
SETUP
YES
°C
ATC
SETUP
P 1.3
°C
ATC
SETUP
°C
HOLD
ENTER
MI
MR
Figure 33: P1.3-
Select Automatic or 
Manual Temperature 
Compensation