8.2.4  P1.4: Selection of TDS Conversion Factor 
You can set the TDS conversion factor between 0.40 and 
1.00; meter default is 0.50.  
NOTE: Your meter has to be in the TDS measurement mode 
before you can make the selection of the TDS Conversion 
Factor in the SETUP menu. 
Only from TDS measurement mode 
1.  Press SETUP key to enter Set Up menu.  
2.  Press the HOLD/ENTER keys until you view 
parameter P1.4.  
3.  Press the HOLD/ENTER key again. The upper 
display shows a value and the lower display shows 
“tdS”. 
4.  Press the MI/▲ or MR/▼ keys to select your 
calculated TDS conversion factor.  
5.  Press the HOLD/ENTER key to confirm selection 
and to return to the subgroup menu. 
6.  Press the CAL/MEAS key to return to measurement 
mode.  
ppt
ppm
SETUP
p 1.4
ppt
ppm
SETUP
0.50
ppt
ppm
SETUP
0.75
MI
MR
HOLD
ENTER
Figure 34: P1.4 -Select 
TDS Conversion 
Factor