9 TROUBLE-SHOOTING 
GUIDE 
Problem Cause 
Solution 
No display 
when turned 
on 
a)  Batteries not in place 
b)  Batteries not in correct 
polarity (+ and – 
position). 
c) Weak 
batteries 
a)  Remove rubber boot/stand. Check 
that batteries are in place and 
making good contact.  
b)  Re-insert batteries with correct 
polarity.  
c) Replace 
batteries. 
 
Unstable 
readings 
a)  Insufficient electrolyte in 
probe.  
b) Air 
bubbles 
trapped 
around the probe. 
c)  Dirty or damaged probe 
d)  Probe not deep enough 
in sample.  
e)  External noise pickup or 
induction caused by 
nearby electric motor.  
f) 
Broken probe.  
a) 
Fill probe with electrolyte & 
replace pre-membraned cap.  
b) 
Stir or tap probe to remove 
bubbles.  
c) 
Clean the probe and re-calibrate.  
d) 
Make sure sample entirely covers 
the probe sensors.  
e) 
Move or switch off interfering 
motor.  
f) 
Replace probe.   
Slow 
response 
a)  Dirty / Oily probe.  
b) Temperature 
is 
changing. 
a) 
Clean probe. See “Probe Care & 
Maintenance. 
b) 
Allow time for temperature to 
stabilize 
Not 
responding 
to key press 
a) 
HOLD mode in 
operation indicated by 
flashing display.  
b) 
Damaged key pad.  
c) 
Internal program error.  
a) 
Cancel HOLD mode.  
b) 
Return to dealer.  
c) 
Reset all internal programs by 
reinserting batteries.