eutech-instruments ecoscan do 6 do meter Manuel D’Utilisation

Page de 76
Instruction Manual 
DO 6 
 
 
57
 
10 ERROR 
MESSAGES 
LCD Display 
Indicates Cause 
Solution 
  
Low 
Battery 
indicator 
Low battery 
level.  
Need new batteries or 
battery connection is 
bad.  
Clean battery contacts. 
Replace batteries with 
fresh ones, noting 
polarity.  
“Err 1” in % 
Saturation Mode 
% Saturation 
Calibration 
error  
Calibration is 
performed when factory 
calibrated absolute 
value is out of 
calibration range – 
10.1% to 49.9% 
Check the value of the 
calibration solution. If 
zero calibration is done, 
make sure the limit of 
10% is not exceeded.  
If message persist, 
recondition your probe.  
“Err. 1” in mg/L 
(ppm) 
Concentration 
Mode 
Concentration 
Calibration 
error  
Calibration is 
performed when the 
calibration solution 
factory absolute 
measurement is below 
2.00 mg/L or ppm 
Check the calibrating 
solution to be above 2 
mg/L or ppm. 
Check that the correct 
temperature and salinity 
setting has been set 
prior calibration. 
“UR”/”OR” with 
blinking “A” 
annunciator in 
Temperature 
Mode 
ATC probe 
error, Under 
Range, Over 
Range 
ATC probe is removed 
or broken while the 
meter has the ATC 
feature activated. 
Check the probe’s 
temperature input 
phono jack connection 
to the meter. Ensure 
probe is not broken or 
punctured. 
“----“with blinking 
“A” annunciator 
in % Saturation 
and 
Concentration 
Mode 
ATC probe 
error 
ATC probe is removed 
or broken while the 
meter has the ATC 
feature activated. 
Check the probe’s 
temperature input 
phono jack connection 
to the meter. Ensure 
probe is not broken or 
punctured.