eutech-instruments sodium glass Manuel D’Utilisation

Page de 16
Sodium Electrode 
 
Instruction Manual 
 
12
persists and you are using an electrode pair, try the same routine with a working reference 
electrode. 
 
4. 
If the problem persists, the reagent may be of poor quality, interference in the sample may 
be present or the technique may be faulty. (See the
 
Standards and ReagentsSample
and
 
Technique 
sections below.) 
 
5. 
If another electrode is not available for test purposes, or if the electrode in use is suspect, 
review the instruction manual and be sure to: 
- Clean and rinse the electrodes thoroughly. 
- Prepare the electrode(s) properly. 
- Use the proper filling solution. 
- Adjust the pH and the ionic strength of the solution by the use of the proper ISA. 
- Measure correctly and accurately. 
- Review
 
TROUBLESHOOTING HINTS
 
Standards and Reagents 
Whenever problems arise with the measuring procedure that has been used successfully in the past, 
be sure to check the standard and reagent solutions.  If in doubt about the credibility of any of the 
solutions, prepare them again.  Errors may result from contamination of the ISA, incorrect dilution 
of standards, poor quality distilled or deionized water, or a simple mathematical miscalculation. 
 
Sample 
Look for possible interference, complexing agents, or substances which could affect the response or 
physically damage the sensing electrode (or the reference electrode) if the electrodes work perfectly 
in the standard, but not in the sample. 
 
Try to determine the composition of the samples prior to testing to eliminate a problem before it 
starts.  (See
 Measuring HintsSample Requirements
and
 
Interference.) 
 
Technique 
Be sure that the electrode's limit of detection has not been exceeded. Be sure that the analysis 
method is clearly understood and is compatible with the sample. 
 
Refer to the instruction manual. Reread sections on 
General Preparation
 
and  Electrode 
Characteristics
. If trouble still persists, please call Eutech Instruments Pte Ltd. at (65) 6778-6876 
and ask for the Customer Services Department.