eutech-instruments sodium glass Manuel D’Utilisation

Page de 16
Instruction Manual 
 
Sodium Electrode 
 
13
TROUBLESHOOTING HINTS 
 
 
 
 
Symptom 
Possible Causes 
Next Step 
Out of Range Reading 
defective meter 
check meter with shorting strap 
(see meter instruction manual) 
 
defective electrode 
check electrode operation 
 
electrodes not plugged in 
properly 
unplug electrodes and reset 
 
reference electrode not filled 
replenish reference filling solution 
 
electrodes not in solution 
put electrodes in solution 
 
air bubble on redipping 
electrode 
remove bubble by membrane 
"Incorrect Answer" (but 
calibration curve is 
good) 
incorrect scaling of semi-log 
paper 
plot millivolts on the linear axis. On 
the log axis, be sure concentration 
numbers within each decade are 
increasing with increasing 
concentration 
 
incorrect sign 
be sure to note sign of millivolt 
number correctly 
 
incorrect standards 
prepare fresh standards 
 
wrong units used 
apply correct conversion factor: 
1.0X10-
3
M = 23.0 ppm as Na
+1
 
 
sample pH too acidic 
adjust pH with ISA 
Drift (reading changing 
in one direction) 
samples and standards at 
different temperature 
allow solutions to come to room 
temperature before measurement 
 
incorrect reference filling 
solution 
use recommended filling solution 
 
electrode exposed to 
interference 
soak overnight in electrode storage 
solution 
 
pH too acidic 
adjust pH with ISA 
 
ISA not used 
use recommended ISA 
Low Slope or No Slope 
standards contaminated or 
incorrectly made 
prepare fresh standards 
 
standard used as ISA 
use ISA 
 
pH too acidic 
adjust pH with ISA 
 
electrode exposed to 
interference 
soak overnight in electrode storage 
solution 
 
air bubble on membrane 
remove bubble by redipping probe 
Noisy or Unstable 
Readings (reading 
changing randomly) 
defective meter 
check meter with shorting strap 
 
ISA not used 
use recommended ISA 
 
meter or stirrer not  grounded 
ground meter or stirrer 
 
defective electrode 
replace electrode 
 
electrode exposed to 
interference 
soak overnight in electrode storage 
solution