eutech-instruments waterproof c401 colorimeter Manuel D’Utilisation

Page de 44
Calibration: Cyanuric Acid
 
 
C401/301/201/102/101 
 
 
 
27
3.3 
Cyanuric Acid 
1. 
Prepare a Cyanuric Acid standard solution in the 66 - 75 ppm range 
and record the concentration value. 
2. 
Ensure the meter is switched off. Press and hold the ZERO key while 
switching the meter on by pressing the ON/OFF key. The meter will 
momentarily flash the model number and then display the CAL 
annunciator at the top left corner of the screen indicating the 
instrument is in calibration mode. 
3. 
Press the MODE key repeatedly until [CYNA] is displayed. 
OFF
ON
ZERO
CAL
MODE
Press Mode to 
select test 
parameter of 
“CYNA” for 
Cyanuric Acid
 
4. 
Press the READ/ENTER key to confirm the selected test. The meter 
will display [bLNK] indicating it is waiting for a 
blanking sample.  
Note: blanking is compulsory during calibration; without 
blanking the meter will not continue with the calibration 
procedure.
  
5. 
Select a clean vial, un-screw the cap and fill with de-ionised water to 
the line on the vial.  Replace the cap, ensuring it is screwed on finger 
tight to the vial. 
6. 
Ensure the outside of the vial is clean, dry and absent of any marks 
or finger-prints. Gently place the vial fully into sample well of the 
instrument, whilst aligning the ▼ mark on the vial with the ▲ mark on 
the meter.  
7.  Press the ZERO key to blank the meter to that vial; the meter will 
display [ZERo] whilst this is taking place. 
8. 
Remove the blanking vial from the meter.  
9. 
The meter displays 66 ppm, with the CAL annunciator on the top left 
corner. 
10.  Use the MODE key repeatedly to select the concentration of the 
prepared calibration standard in the 66 - 75 ppm range. 
Blanking 
Sample
CAL
MODE
ZERO
CAL
ppm
Press Mode to 
select calibration 
point from 66 
ppm to 75 ppm in 
an incremental 
circular manner.
 
CAL