eutech-instruments waterproof c401 colorimeter Manuel D’Utilisation

Page de 44
Calibration: Cyanuric Acid
 
 
C401/301/201/102/101 
 
 
 
28
Shake for
1 minute
11.  Select another clean vial, un-screw the cap and fill it with 10 ml of the 
prepared Cyanuric Acid standard solution (fill up to the line on the 
vial).   
12.  Select a Cyanuric Acid sachet. Open the sachet and tip the contents 
into the vial containing the calibration standard. 
13.  Replace the cap, ensuring it is screwed on finger tight 
to the vial.  Shake the vial for approximately one 
minute, until the powder has dissolved.  
Note: a small amount of the powder may remain un-dissolved in the vial, but 
this will not affect the measurement. 
14.  Insert the vial fully into the sample well of the meter, whilst again 
aligning the ▼ mark on the vial with the ▲ mark on the meter. 
15.  Wait for two minutes, then press the READ/ENTER key within five 
minutes from the time the reagent was added to the vial. 
16.  The meter will display [--Rd--] whilst the calibration standard is being 
measured and then displays [SAVE] to prompt for calibration 
acceptance and saving of calibration data. 
Cyanuric
   Acid 
Sample 
  70 ppm
ppm
READ
ENTER
CAL
 
17.  Press the READ/ENTER key. The meter will now accepts the 
calibration value and momentarily display [HoLD] and CAL 
annunciator before returning to the beginning of the calibration as in 
step 4. 
18.  To continue with another calibration, press the MODE key to select 
the desired calibration mode. 
19.  To exit the calibration mode, press the ZERO key to return to the 
measurement mode. 
CAL
MODE
READ
ENTER
Press Mode to 
select another 
calibration test 
ZERO
Press ZERO to 
return to 
measurement 
mode
CAL