6 CONDUCTIVITY 
CALIBRATION 
This section is applicable for the following models:  
Description 
Eutech Instruments 
Oakton Instruments 
Conductivity FMCON 1 with Conductivity 
electrode and 110/120 VAC power adapter. 
EC-CONFM1ST11 35660-00 
Conductivity FMCON 1 with Conductivity 
electrode and 220/230 VAC power adapter. 
EC-CONFM1ST21 35660-05 
Conductivity FMCON 2 with Conductivity 
electrode and 110/120 VAC power adapter. 
EC-CONFM2ST11 35660-02~ 
Conductivity FMCON 2 with Conductivity 
electrode and 220/230 VAC power adapter. 
EC-CONFM2ST21 35660-07~ 
Conductivity FMCON 3 with Conductivity 
electrode and 110/120 VAC power adapter. 
EC-CONFM3ST11 35660-04~ 
Conductivity FMCON 3 with Conductivity 
electrode and 220/230 VAC power adapter. 
EC-CONFM3ST21 35660-09~ 
6.1  Important Information on Conductivity/TDS Monitor Calibration 
This monitor has ±30% adjustment. Adjustable in 0.1 mS increments. Choose a 
calibration standard that is close to your working solution.  
Calibration should be performed regularly to ensure accuracy. Frequency of 
calibration will vary and should be checked regularly until the required schedule has 
been determined.  
Do not reuse calibration solutions after calibration. Dispose of any used calibration 
solution. Contaminants in the solution can affect the calibration, and eventually the 
accuracy of the measurements.  
NOTE: All new calibration data will over-ride existing calibration data. Minimum 
calibration value is 2.5 mS.  
Adjustments and Minimum Calibration Value are according to specific conductivity 
meter model: 
 
Models 
Calibration Adjustments 
Minimum Calibration Value 
FM CON 1 
0.1 mS increments 
2.5 mS 
FM CON 2 
10 
µS increments 
250 
µS 
FM CON 3 
µS increments 
37.5 
µS