5 ORP 
CALIBRATION 
This section is applicable for the following models:  
•  Eutech Instruments: EC-ORPFM1ST11 & EC-ORPFM1ST21 
•  Oakton Instruments: 35649-00 and -05 only 
5.1  Important Information on ORP Monitor Calibration 
Typically ORP measurements are taken on a relative basis and calibration is not 
required. This monitor has ±100 mV offset adjustment, adjustable in 5 mV 
increments. Choose a calibration standard that is close to your working condition.  
Calibration should be performed regularly to ensure accuracy. Frequency of 
calibration will vary and should be checked regularly until the required schedule has 
been determined.  
Be sure to remove the protective electrode storage bottle or rubber cap of the probe 
before calibration or measurement. If the electrode has been stored dry, wet the 
probe in tap water for 10 minutes before calibrating or taking readings to saturate the 
ORP electrode surface and minimize drift.  
Do not reuse calibration solutions after calibration. Dispose of any used calibration 
solution. Contaminants in the solution can affect the calibration, and eventually the 
accuracy of the measurements.  
IMPORTANT: ORP offset adjustment is limited to ±100 mV from the original reading. 
Adjustable in 5 mV increments.  
NOTE: All new calibration data will over-ride existing calibration data.