Ophir Optronics Ltd QSR Manuel D’Utilisation

Page de 120
Chapter 1 – Introduction 
 Technical and System Performance Specifications 
Technical and System Performance Specifications 
USBI and Pulsar Specifications 
The USBI and Pulsar technical and system performance specifications are 
described in Table 1-1
Table 1-1 USBI and Pulsar Technical and System Performance Specifications 
Item 
USBI Specification 
Pulsar Specifications 
Specifications for Thermopile and Photodiode heads: 
Measurement range 
Varies according to head in use.  
Refer to the Ophir Laser Power/Energy Measurement, at 
www.ophiropt.com, for full details of each individual head. 
 
 
 
Input range 
15nA to 1.5mA in 16 ranges  15nA to 1.5mA in 16 ranges 
A-to-D sampling rate
15Hz 
15Hz 
A-to-D resolution 
17 bits plus sign (0.0009% 
resolution) 
17 bits plus sign 
Electrical accuracy 
±0.25% ±20pA new;  
±0.5% ±50pA after 1 year 
±0.25% ±20pA new;  
±0.5% ±50pA after 1 year 
Electrical input noise 
level 
500nV or 1.5pA +0.0015% 
of input range @3Hz 
500nV or 1.5pA +0.0015% of input 
range @3Hz 
Dynamic range 
9 decades (1:10^9) 
9 decades (1:10^9) 
Specifications for Pyroelectric and PD10 heads: 
Measurement range 
Varies according to head in use.  
Refer to the Ophir Laser Power/Energy Measurement, at 
www.ophiropt.com, for full details of each individual head. 
Input range 
0 – 6v full scale 
0 – 6v full scale 
A-to-D sampling 
4kHz 
>20kHz 
A-to-D resolution 
12 bits no sign (0.025%) 
12 bits no sign (0.025%) 
Electrical accuracy 
±0.25% new, ±0.5% after 1 
year 
±0.25% new, ±0.5% after 1 year 
Electrical input noise  2mV 
2mV 
Logging rates, Turbo 
mode 
2kHz every pulse 
20kHz, every pulse on all 4 channels 
[using USB 2.0] 
Logging rates, 
normal log mode 
~ 300Hz 
~ 1kHz [using USB 2.0; varies according 
to PC's CPU speed and number of 
channels running] 
Log file timestamp 
precision 
1ms (0.001s) 
1us (0.001ms) 
Log file timestamp 
resolution 
~50ms 1us 
(0.001ms) 
StarLab User Guide