3.4 
Generator Interface 
VIVIX-S 1012N detector provides DR Trigger interface and AED interface as a generator interface way to 
acquire images by detecting X-ray. 
 
Mode 
Description 
AED 
The detector detects X-ray exposure from the generator automatically and then performs 
image acquisition without any cable connection. 
DR Trigger 
The detector and generator receive and send their signal to each other for image 
acquisition. SCU and X-ray generator should be connected with the generator interface 
cable. 
 
3.4.1  AED (Auto Exposure Detection) Interface 
If the VIVIX-S 1012N detector is used as the AED interface, you can acquire images without connecting the 
generator to the detector with a generator interface cable. 
 
            
 
 
 
Cautions when using AED interface 
 
 
Make sure to follow the operating environmental condition (Temp: +10℃  ~ +35℃). 
 
Do not give impact to the product. If it receives strong jolt, unwanted images may be 
acquired without the X-ray exposure because of the malfunction of the AED sensor. 
 
You may not acquire images or horizontal artifacts may occur depending on the external 
environment such as exposure condition, thickness of object or the use of grid. 
 
When you set X-ray exposure area to the direction of the detector, the center of the 
detector should be involved in the X-ray exposure area. Otherwise, you may not acquire 
an image.