Generator Interface
VIVIX-S 1012N detector provides DR Trigger interface and AED interface as a generator interface way to
acquire images by detecting X-ray.
acquire images by detecting X-ray.
Mode
Description
AED
The detector detects X-ray exposure from the generator automatically and then performs
image acquisition without any cable connection.
image acquisition without any cable connection.
DR Trigger
The detector and generator receive and send their signal to each other for image
acquisition. SCU and X-ray generator should be connected with the generator interface
cable.
acquisition. SCU and X-ray generator should be connected with the generator interface
cable.
3.4.1 AED (Auto Exposure Detection) Interface
If the VIVIX-S 1012N detector is used as the AED interface, you can acquire images without connecting the
generator to the detector with a generator interface cable.
generator to the detector with a generator interface cable.
Cautions when using AED interface
Make sure to follow the operating environmental condition (Temp: +10℃ ~ +35℃).
Do not give impact to the product. If it receives strong jolt, unwanted images may be
acquired without the X-ray exposure because of the malfunction of the AED sensor.
acquired without the X-ray exposure because of the malfunction of the AED sensor.
You may not acquire images or horizontal artifacts may occur depending on the external
environment such as exposure condition, thickness of object or the use of grid.
environment such as exposure condition, thickness of object or the use of grid.
When you set X-ray exposure area to the direction of the detector, the center of the
detector should be involved in the X-ray exposure area. Otherwise, you may not acquire
an image.
detector should be involved in the X-ray exposure area. Otherwise, you may not acquire
an image.