3.4.2  DR Trigger Interface 
Connect the detector and X-ray generator with a generator interface cable, and then acquire images by 
sending and receiving their signal. 
 
 
 
Signal Processing Steps 
1  The detector receives EXP_REQ signal from the X-ray generator. 
2  After the detector completes to prepare image acquisition, it sends the EXP_OK signal to the X-ray 
generator. 
3  The X-ray generator confirms EXP_OK signal and generates X-rays. 
4  The detector acquires images and then transmits the image data. 
 
 
 
EXP_REQ is a sending signal from X-ray generator to the detector for requesting 
exposure. 
 
EXP_OK is an exposure ready signal from the detector to X-ray generator.