Vieworks Co. Ltd. FXRD-1012NAW Manuel D’Utilisation

Page de 151
 
 
Rev.1.0
 
Page 85 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
 
Trigger 
Configures trigger information for integration between SCU and X-ray generator. 
Item 
Description 
Method 
Trigger mode 
Packet 
Software Trigger mode 
Line 
Hardware Trigger mode 
Polarity 
Polarity of trigger signal 
Auto (Default) 
Recognizes polarity automatically and handles it. (Default value) 
Active High 
Handles polarity of Active High 
Active Low 
Handles polarity of Active Low 
 
 
 
Configuration value of trigger is applied only when Exposure mode is set DR Trigger
Configuration value of trigger is not applied when using AED mode. 
 
 
 
The setting of Trigger should be done by an engineer who understands about the x-ray 
generator device well. Unless the device and detector are set correctly, an integration 
error of x-ray generator would occur or the system operation would be affected. 
 
 
Command Buttons 
Item 
Description 
Set Config 
Updates by transmitting current setting values to SCU. 
Factory Reset 
Resets SCU to factory default settings. 
Log 
Checks the logs of SCU. 
Close 
Closes the window of SCU Configuration. If Set Config is not performed, the 
changed setting value is not transmitted to SCU.