Trigger
Configures trigger information for integration between SCU and X-ray generator.
Item
Description
Method
Trigger mode
Packet
Software Trigger mode
Line
Hardware Trigger mode
Polarity
Polarity of trigger signal
Auto (Default)
Recognizes polarity automatically and handles it. (Default value)
Active High
Handles polarity of Active High
Active Low
Handles polarity of Active Low
Configuration value of trigger is applied only when Exposure mode is set DR Trigger.
Configuration value of trigger is not applied when using AED mode.
Configuration value of trigger is not applied when using AED mode.
The setting of Trigger should be done by an engineer who understands about the x-ray
generator device well. Unless the device and detector are set correctly, an integration
error of x-ray generator would occur or the system operation would be affected.
generator device well. Unless the device and detector are set correctly, an integration
error of x-ray generator would occur or the system operation would be affected.
Command Buttons
Item
Description
Set Config
Updates by transmitting current setting values to SCU.
Factory Reset
Resets SCU to factory default settings.
Log
Checks the logs of SCU.
Close
Closes the window of SCU Configuration. If Set Config is not performed, the
changed setting value is not transmitted to SCU.
changed setting value is not transmitted to SCU.