VIVIX-S 1012N User Manual 
6.1 
Calibration Dialogue 
Use the following menus to configure the system and process calibrations. 
Menu 
Description 
System Configuration 
Configures the exposure mode, exposure-related time and Gain type. 
Offset Calibration 
Processes Offset calibration. 
Defect Calibration 
Processes calibration for defect data of the detector. 
Gain Calibration 
Processes calibration for image sensitivity. 
Detector Configuration 
Configures the detector information related to calibration and image. 
 
 
 
The menu status is displayed as Done when each calibration is finished. You cannot 
operate the detector if the status is not displayed as Done
6.1.1  System Configuration Dialogue 
Configure the generator interface information from System Configuration in Step
 
 
Exposure Mode 
The VIVIX-S 1012N detector provides two types of mode for taking images as follows. 
Mode 
Description 
AED   
Detects X-ray automatically without connecting the detector to X-ray generator. 
The detector acquires images after the automatic detection. 
DR Trigger 
Detects the X-ray exposure signal of X-ray generator in advance by connecting the 
detector to X-ray generator with a generator interface cable. 
The detector acquires images after preparing exposure with the exposure signal.