Exposure Timing
To take images exactly, set the timing information for exposure.
The information can be different depening on the characteristics of the X-ray generator.
Exposure Timing Diagram
Exposure Timing Items
Item
Default Description
Exposure section
500㎳
Time to acquire X-ray by the detector.
Pre Exposure section
0㎳
Standby time after the detector sent EXP_OK signal to the generator.
EXP OK Delay section 1㎳
The delayed time to receive EXP_OK signal from the detector after the
generator sent EXP_REQ signal to the detector.
generator sent EXP_REQ signal to the detector.
Debounce_DR
3㎳
The required time to check signals from the generator for preventing
image acquisition from external noise. It can be used when DR Trigger
mode is selected only.
image acquisition from external noise. It can be used when DR Trigger
mode is selected only.
Debounce_AED
3㎳
The required time to check sensor signal from X-ray to prevent image
acquisition from external noise. It can be used when AED mode is
selected only.
acquisition from external noise. It can be used when AED mode is
selected only.
Exposure section should be set longer than the maximum exposure time to prevent X-ray
loss.
loss.