6.1.2 Offset Calibration Dialogue
Vieworks provides the Defect Map data stored in the detector. However, we recommend
you to carry out the Offset calibration by yourself as the detector condition can be
different by the operating method or use environment.
you to carry out the Offset calibration by yourself as the detector condition can be
different by the operating method or use environment.
Offset Calibration
Item
Description
Current value
Shows value when performing the manual Offset calibration.
Stage
Selects the performance frequency of manual Offset calibration and
shows each calibration stage.
shows each calibration stage.
Run Pre-offset Calibration
Performs the manual Offset calibration.
Load Pre-offset
Downloads the Offset data stored in the detector.
Offset Path
Shows the path of the applied Offset data.