Monitor Setup
M310 Fall Monitor – Setup & User Guide 13
4. 
Press the monitor’s Reset Button to confirm that the status LED 
flashes green, and ONLY Green.
CAUTION: 
If a yellow flash is seen, diagnose the problem using 
the LED and Audible Tones table on page 7.
5. 
Replace monitor batteries if a low battery is indicated.
6. 
Apply FULL and FIRM PALM pressure with your hand to the Pad 
for more than 3 seconds.
7. 
The Monitor should flash its Status LED green for 3 seconds and 
then sound a low-high confirmation tone followed by a long 
green flash.
8. 
Release hand pressure. After a delay up to 4.75 seconds 
(depending on configured delay settings), the Monitor should 
generate a Pad Exit alarm, continuously flashing the Alarm LED 
red and, unless a Nurse Call system is connected and the volume 
is set to NC, sounding the Pad Exit Alarm.
9. 
Press the Reset button on the Monitor to silence the alarm. 
10. 
If the Monitor is used with a Nurse Call system, verify that an 
alarm was triggered on that system.
11. 
If the Monitor did not signal a Pad Exit alarm properly, try 
checking all connections, then re-connect the pad and repeat all 
the above tests. 
12. 
If the Pad Exit alarm still does not sound, DO NOT place the pad 
or the monitor into service.
Changing the Sensormat Pad
To change the Sensormat pad while monitoring a patient:
1. 
Put the monitor in Hold Mode, and remove the patient/resident 
from the bed/chair. (See “Hold Mode” on page 9.)
2. 
Remove the old pad from the bed or chair.
3. 
Unplug the old pad from the monitor.
4. 
Replace the pad.
5. 
Plug in the new pad.
6. 
Verify the proper installation by following the instructions on 
the pad.
7. 
Test the new pad’s functionality.