Monitor Setup
M310 Fall Monitor – Setup & User Guide 11
To Insert Batteries
1. 
Insert the batteries into the battery compartment, by pressing 
the negative end of each battery into the corresponding spring, 
then sliding the positive end into the contact. 
The monitor immediately performs a self-test. The LEDs flash in 
all colors and the speaker sounds.
2. 
Replace the battery cover, and insert and tighten the cover 
screw.
External Power
External power may be used but the batteries must be installed 
for back-up power. When external power is used and batteries are 
installed power is drawn only from the external power.
The external power adaptor used must supply 5V/1A only. Currently, 
model FRM015-505-x, (sold by STANLEY) is the only approved power 
supply for the M310.
Using Sensormat Pads
Locate the Sensormat pad either on a bed or chair, directly underneath 
the patient/resident.
Either 14-day or 6-month (180-day) pads can be deployed with the 
monitor.
Note: 
Pads should be replaced on or before their warranty 
period expiry date (indicated on pad).
Sensormat pads are designed to sense body-weight distributed over an 
area.
Pad testing should be done either by sitting or lying on the pad, or by 
pressing down firmly with the palm of your hand.
For more information on pad testing see “Testing the Sensormat Pad” 
on page 12.
On a Bed
Place the Sensormat pad across the width of the bed, ON TOP OF THE 
MATTRESS.
A top sheet and/or incontinence pad may be placed above the 
Sensormat pad.