Options
With the Precision SCS system, the Pulse Width and Rate parameters
are generally considered programming “options” in that relatively few
patients will actually use these settings. By default, Pulse Width and
Rate are off-limits to the patient Remote Control unless you allow
access using BionicNavigator software.
With the Precision SCS system, the Pulse Width and Rate parameters
are generally considered programming “options” in that relatively few
patients will actually use these settings. By default, Pulse Width and
Rate are off-limits to the patient Remote Control unless you allow
access using BionicNavigator software.
Another option, “Restore,” is readily available to all patients. The
Restore feature allows patients to return a program to the original set-
tings you programmed for them at the initial fitting or at a follow-up.
Restore feature allows patients to return a program to the original set-
tings you programmed for them at the initial fitting or at a follow-up.
To Access the Pulse Width and Rate Options:
1. Press and hold the
C
button for three seconds (a long button
press) or until the Pulse Width Area 1 screen appears.
2. As necessary, press the
C
button normally to cycle to the area
you wish to adjust.
3. When the desired area screen is displayed, press S or T to
increase or decrease the Pulse Width.