Rehabtronics Inc. HSST1 Manuel D’Utilisation

Page de 52
34 
 
Adjusting Intensities 
 
 
IMPORTANT: If sharing the STIM between users, intensities should always be reset 
between users prior to following these instructions (see previous section). 
 
NOTE: Increasing intensities requires the STIM to be connected to a GARMENT, allowing the person 
wearing the GARMENT to feel the effects as the intensity is adjusted. 
 
Stimulation intensity can be adjusted individually to the HAND-OPEN, HAND-CLOSE and THUMB 
ELECTRODES as follows: 
 
1.  Wet the ELECTRODES and fit the GARMENT according to the instructions in the previous 
sections. 
2.  Insert the STIM into the GARMENT’s STIM-HOLDER
3.  Turn the STIM on by pressing and holding the ON/OFF BUTTON
4.  Press the 
button to arm the STIM
5.  Select “STIM INTENSITY” from the Main Menu. 
6.  Use the 
button to access the “HAND OPEN”, “HAND CLOSE” and “THUMB CLOSE” 
screens. 
7.  Use the  
  button to gradually increase the intensity of stimulation to the each electrode, 
until the desired hand-response is achieved. 
 
CAUTION: Do not increase to painful levels.  The desired hand-response should be 
achievable without pain or discomfort.  If significant discomfort is experienced without 
achieving the desired response, reposition the electrode to another location in the 
PLACEMENT AREAS and try again. 
 
 
NOTE: In most cases, effects will not be felt until the setting has increased beyond 5. 
NOTE: Stimulation can be stopped at any time by pressing the 
button. 
NOTE: Not all circumstances require thumb stimulation.  If thumb stimulation is not needed, the 
THUMB-CLOSE screen can be left at a setting of ‘OFF’.