Controlling Stimulation by Tapping 
Functional stimulation control by tapping the STIM is achieved as follows: 
 
1. Check that there is no HMS connected to the STIM.  The top right of the 
DISPLAY’s Status Bar should show “TAP”. 
If “TAP” is not shown in the Status Bar, check that any nearby HMS unit has 
been turned off. 
[image] 
2. Press the 
button to arm the STIM.  The Stimulation Status icon at the center 
of the DISPLAY’s Status Bar should be solid gray 
 
3. Select “FES MODE” from the Main Menu. 
[image] 
4. Tap the STIM to cycle between hand-opening, hand-closing and relax (no 
stimulation). 
NOTE: Stimulation can be stopped at any time by pressing the 
button. 
 
Adjusting the Tap Sensitivity 
To adjust how sensitive the STIM is to taps, do the following: 
 
1. Select “DEVICE SETUP” from the STIM’s Main Menu. 
2. Press the 
button until the “TAP SENSITIVITY” screen is reached. 
 
3. Use the 
and 
 buttons to adjust the tap sensitivity: 
Press the 
 button to make the STIM less sensitive (harder tap required) 
Press the 
button to make the STIM more sensitive (softer tap required) 
 
NOTE: The current setting can be tested by tapping the STIM.  The dot on the screen will flash green if 
the tap was hard enough for the STIM to detect it at the current sensitivity setting.