Rehabtronics Inc. HSST1 Manuel D’Utilisation
36
Controlling Stimulation by Tapping
Functional stimulation control by tapping the STIM is achieved as follows:
1. Check that there is no HMS connected to the STIM. The top right of the
DISPLAY’s Status Bar should show “TAP”.
DISPLAY’s Status Bar should show “TAP”.
If “TAP” is not shown in the Status Bar, check that any nearby HMS unit has
been turned off.
been turned off.
[image]
2. Press the
button to arm the STIM. The Stimulation Status icon at the center
of the DISPLAY’s Status Bar should be solid gray
.
3. Select “FES MODE” from the Main Menu.
[image]
4. Tap the STIM to cycle between hand-opening, hand-closing and relax (no
stimulation).
stimulation).
NOTE: Stimulation can be stopped at any time by pressing the
button.
Adjusting the Tap Sensitivity
To adjust how sensitive the STIM is to taps, do the following:
1. Select “DEVICE SETUP” from the STIM’s Main Menu.
2. Press the
button until the “TAP SENSITIVITY” screen is reached.
3. Use the
and
buttons to adjust the tap sensitivity:
Press the
button to make the STIM less sensitive (harder tap required)
Press the
button to make the STIM more sensitive (softer tap required)
NOTE: The current setting can be tested by tapping the STIM. The dot on the screen will flash green if
the tap was hard enough for the STIM to detect it at the current sensitivity setting.
the tap was hard enough for the STIM to detect it at the current sensitivity setting.