System Cleaning and Maintenance
General System Maintenance
The following general maintenance guidelines should be followed to ensure safe and effective use of the
Hand Stimulator system:
Hand Stimulator system:
Inspect all components prior to use. Do not use any components show signs of damage.
Do not store or operate any of the system components outside their specified operating and
storage conditions (see “General Specifications” section of this manual).
storage conditions (see “General Specifications” section of this manual).
All system components can be cleaned of surface debris as necessary by wiping with a soft cloth
dampened with clean water.
dampened with clean water.
WARNING: Do not perform any cleaning or maintenance while the equipment is in use.
Ensure components are powered off prior to following any of the cleaning instructions in
this section.
Ensure components are powered off prior to following any of the cleaning instructions in
this section.
DOCK Cleaning & Maintenance
Cleaning the DOCK can be performed as necessary to remove surface debris and prevent obstruction of
the STIM CRADLE and HMS CRADLE. If there is noticeable dirt, dust or other substance accumulating on
the surface or within either of the STIM or HMS cradles, perform the following:
the STIM CRADLE and HMS CRADLE. If there is noticeable dirt, dust or other substance accumulating on
the surface or within either of the STIM or HMS cradles, perform the following:
1. Unplug the DOCK.
2. Wipe the DOCK’s outer surfaces with a soft cloth dampened with clean water.
3. Use compressed air to dislodge any obstructions from the STIM and HMS cradles.
STIM and HMS Cleaning & Maintenance
The STIM and HMS can be cleaned, as needed, by simply wiping with a soft cloth dampened with clean,
warm water.
warm water.
If low-level disinfecting is required, such as in clinical settings, the STIM and HMS can be wiped with a
cloth soaked with 70% isopropyl alcohol (IPA) solution. Continue to wipe as necessary to ensure all
surfaces are kept wet with the solution for at least three (3) minutes.
cloth soaked with 70% isopropyl alcohol (IPA) solution. Continue to wipe as necessary to ensure all
surfaces are kept wet with the solution for at least three (3) minutes.
WARNING: Do not immerse the STIM or HMS in water or any other liquid. Doing so can
permanently damage the electronics.
permanently damage the electronics.
When not in use, it is recommended that the STIM and HMS be stored in the DOCK. This will prolong the
battery life, and ensure that the STIM and HMS are fully charged when next used.