Sutron Corporation RLR-0003-1 Manuel D’Utilisation

Page de 63
Bringing the Benefits of Real-Time Data Collection to the World 
14 
Stage, Distance to Water, and Measurement Point 
Ultimately, the job the radar is to compute a stage measurement.  However, the direct measurement made by 
the radar is distance to water.   
 
Distance to water is the length between the radar and the water surface.  
 
Stage is the depth of water; for a river it is distance between the water surface and the riverbed.   
 
Stage = Measurement point – Distance to water 
 
Measurement point is the length between the radar and the riverbed (or ocean floor or well bottom). 
(NOTE:  Calibration  of  the  radar  unit  has  been  made  with  respect  to  the  front  cover  of  the 
antenna unit.) 
 
 
Measuring Point: 
Bridge or  Structure over Water 
Stage 
Water Level (Stage) 
Riverbed: 
Distance to Water 
Staff 
Gauge 
Stage Reading: 
Measuring Point = Stage + Distance to Water 
 
 
 
or 
Stage = Measuring Point – Distance to Water 
 
 
 
Stage is displayed on the first menu shown on th
.  Via command line, use the 
” command 
or the 
” command.  Via SDI-12, use th
. 
 
Measurement point can be accessed via the front panel 
, via SDI-12 using th
, and the command lin
. 
 
Setting Stage 
When the radar is installed, you will typically read the current  stage off a staff gauge and then set this new 
stage into the radar.   
 
To set the stage via front panel, power up the radar, wait for it to show a stage reading (which will be negative 
until the stage is set), and the press SET.  Enter the correct stage  and press set again.  Don‟t forget to change 
the sign of the stage to “+
”.   It is also possible to use SDI-12 
 and command lin
 to set 
the stage. 
 
The radar compares this user entered value with its own reading and uses it to set the Measuring Point.  From 
then on, the radar will report the stage relative to this Measuring Point – so it will match the staff gauge.