3 Channel (5 lead) Electrode Placement (1
st
option)
Five color-coded leadwires are used to create a 3-channel ECG
recording. This is a typical electrode placement. Refer to your Analysis
System software and the physician for a recommended position.
recording. This is a typical electrode placement. Refer to your Analysis
System software and the physician for a recommended position.
5 Lead Electrode Placement
#
Channel
Color
Location
1
3-
White
Next to the right Manubrium border on the
Clavicle
Clavicle
2
1-, 2-
Red
Centered on the Manubrium
3
2+, 3+
Black
Lower left rib margin over bone.
4
1+
Brown
Left Anterior Auxiliary line on the 6
th
rib
5
Green
Lower right rib margin over bone.
Fusion Wireless Recorder
10
3 Channel (5 lead) Electrode Placement (2
nd
option)
Five color-coded leadwires are used to create a 3-channel ECG
recording. This is a typical electrode placement. Refer to your Analysis
System software and the physician for a recommended position.
recording. This is a typical electrode placement. Refer to your Analysis
System software and the physician for a recommended position.
Figure 4 - 5 Lead Electrode Placement
#
Channel
Color
Placement
1
3-
White
Right side below the V1 position, at the
bottom of the rib cage
bottom of the rib cage
2
1-, 2-
Red
Center on the Manubrium, the top of the
sternum
sternum
3
2+, 3+
Black
Left side at the V5 position, on a rib
4
1+
Brown
Left side at the V3 position, on a rib
5
Green
Right side opposite V5 position.