Braemar Manufacturing LLC FUSIONMCT Manual Do Utilizador

Página de 17
 
 
 
 
      
    Fusion Wireless Recorder 
 
3 Channel (5 lead) Electrode Placement (1
st
 option) 
Five color-coded leadwires are used to create a 3-channel ECG 
recording. This is a typical electrode placement.  Refer to your Analysis 
System software and the physician for a recommended position.   
 
 
 
 
 
 5 Lead Electrode Placement 
 
Channel 
Color 
Location 
3- 
White 
Next to the right Manubrium border on the 
Clavicle 
1-, 2- 
Red 
Centered on the Manubrium 
2+, 3+ 
Black 
Lower left rib margin over bone. 
1+ 
Brown 
Left Anterior Auxiliary line on the 6
th
 rib 
 
Green 
Lower right rib margin over bone. 
 
 
 
 
      
    Fusion Wireless Recorder 
 
10 
3 Channel (5 lead) Electrode Placement (2
nd
 option) 
Five color-coded leadwires are used to create a 3-channel ECG 
recording. This is a typical electrode placement.  Refer to your Analysis 
System software and the physician for a recommended position. 
 
 
Figure 4 - 5 Lead Electrode Placement 
 
Channel 
Color 
Placement 
3- 
White 
Right side below the V1 position, at the 
bottom of the rib cage 
1-, 2- 
Red 
Center on the Manubrium, the top of the 
sternum 
2+, 3+ 
Black 
Left side at the V5 position, on a rib 
1+ 
Brown 
Left side at the V3 position, on a rib 
 
Green 
Right side opposite V5 position.