Table 1.  Key Functions
Left side of key pad:
0 (stop), 10, 25, 50, 
and 100 mm/sec 
(paper-speed keys)
Press the speed keys to specify the paper speed for 
the internal printer/recorder. The printout will show 
the date and time, lead(s) being printed, gain setting, 
chart speed, and filter setting. To stop the printer/
recorder, press the speed key labeled “0” (zero).
Press the paper-feed key to scroll the printer paper 
on the internal printer/recorder.
Press this key to cause the internal printer/recorder 
to print a one-mV calibration pulse.
Press this key to force the trace back to the baseline 
after a defibrillation shock.
Right side of key pad
Press STAT PACE when in telemetry communication 
with the pulse generator to program predetermined 
parameters for emergency bradycardia pacing.
Press DIVERT THERAPY when in telemetry 
communication with the pulse generator to divert 
tachycardia therapy delivery.
Press STAT SHOCK when in telemetry 
communication with the pulse generator to program 
predetermined high-energy parameters for 
emergency shock delivery.
STAT PACE
DIVERT THERAPY
STAT SHOCK